|
1.5.2.1
|
|
1.5.2.2
|
|
•
|
Choosing the ‘All three’ option results in a more complex table format arrangement where each Sample-Variant cell gets subdivided into three sub-cells: the two side-by-side orientation-specific sub-cells are surmounted by the third, ‘Combined’ sub-cell (see Figure 1‑48, above). Again, a small down-pointing triangle appears to the left of the combined value if the AVA software detects a significant difference between the combined Variant frequency value and that of either orientation.
|
|
1.5.2.3
|
|
•
|
Clicking the ‘Change min/max %’ button opens the ‘Set min / max filter %’ window where you can set the minimum and maximum Variant frequency values (percentage) to use as filters, to 2 decimal places (Figure 1‑54). Values must be within the range ‘0.00’ to ‘100.00’, and the maximum value must be greater than or equal to the minimum value. Click ‘OK’ to accept and apply your min/max filter selections. The ‘All’ button resets the values to a minimum of 0.00 and a maximum of 100.00 (thus All Variant frequency values pass filter).
|
|
◦
|
‘Forward or reverse’ causes the min/max settings to be applied only to the orientation-specific Variant frequency values. Either the forward or the reverse Variant frequency (or both) must meet both the minimum and the maximum filters for the Sample-Variant cell to survive the filters and remain in the table as a cell with a white background (rather than being grayed-out for failing the filter).
|
|
◦
|
‘Forward and reverse’ requires that the forward and reverse Variant frequency values both meet the minimum and maximum filters independently. If one orientation fails, the cell does not survive the filter and is grayed-out.
|
|
1.5.2.4
|
|
1.5.2.5
|
|
1.5.2.7
|