skip to main content
Roche logo
1.5.1.1
General Organization
The Variants Frequency Table shows results one Variant per row and one Sample per column (Figure 1‑47). Initially, cells that contain data are white and cells that contain no data are grayed-out (this can happen, for example, if a Sample has no associated Amplicons whose sequence covers the Variant; see below). When an entire row or column is grayed-out, it is moved to the bottom or right of the Table, respectively. This grayed-out scheme is also used by various display features whereby you can filter out the data according to certain criteria, leaving the data of most interest in white cells, in the upper-left area of the Table. (The ‘Compact table’ option from the Variant data display controls can then remove from view all completely grayed-out rows and columns; see section 1.5.2.4)
While the first three columns are always visible, you can scroll through the Sample columns using the scroll bar located at the bottom-right of the Table [see Figure 1‑48, which displays the same data as in Figure 1‑47 but in a more expanded display (see section 1.5.2), showing the scroll bar]. As you scroll to the right, the leftmost Sample columns appear to slide behind the first three rows, so you may end up in situations where you display a partial column just after the ‘Max’ column.
Figure 1‑48: The Variants Frequency Table showing the same data as in Figure 1‑47, but in a more expanded form, showing the scrolling feature that applies to the ‘Sample’ columns.
1.5.1.2
Organizing Data in the Variants Frequency Table
If you right-click on a cell in the columns underneath the Variant or Max headers or if you right-click on a Sample/Variant intersection cell, there is also a “Define Haplotype” option (Figure 1-49 F and G). This option is inactive unless rows for two or more Variants from the same Reference are selected. The “Define Haplotype” option is described in section 1.5.1.4.
1.5.1.2.1
Sort options
1.5.1.2.2
Ignore filters
As indicated above (section 1.5.1.1), the software grays-out cells that contain no data and shifts rows and columns that contain only gray cells to the bottom or right ends of the Table. This focuses the cells of interest (white) to the upper-left area of the Table. With the ‘always ignore column/row’ options, you can gray-out any columns/rows (which moves them to the right/bottom of the Table) to help focus on data of current interest. If you want to gray-out most columns/rows, you can use the ‘ignore all’ option to gray them all first, and then apply the ‘show’ filter (see below) to re-focus on only the ones for which you have a current interest. (Cells will also be grayed-out if they fail the Min/Max filter, from the Variant data display controls; see section 1.5.2.3.)
1.5.1.2.3
Show filters
1.5.1.2.4
Option reversions
If you want to undo all the sorts and filters you applied (i.e. restore all defaults), use the ‘Reset table’ button.
The ‘Reset table’ button removes all sorting and ignore/show filters that have been applied to the table and restores the table to its state prior to the filters. Note that this does not affect any of the table formatting options from the Variant data display controls (see section 1.5.2), including the Min/Max filter settings (section 1.5.2.3).
1.5.1.3
Populating the Global Align Tab from the Variants Tab
1.5.1.4
Defining a Haplotype from the Variants Tab
Right-clicking on a cell in the columns underneath the Variant or Max headers, or right-clicking on a Sample/Variant intersection cell, reveals a “Define Haplotype” option (Figure 1-49 F and G) in the contextual menu. This option uses a multi-row selection of individual Variants in the Variant Frequencies Table to propose a new Variant that requires that all the selected Variants be found together as a haplotype. This option is grayed-out and inactive unless rows for two or more Variants from the same Reference are selected (a multi-selection of Variant rows from mixed References will inactivate the option even if any of the selected Variants share a Reference). After selecting two or more such Variant rows, right-clicking over any cell in the selection except for a Reference cell provides access to an active “Define Haplotype” option (Figure 1‑51).
Choosing the option will bring up the “Define Haplotype” window (Figure 1‑52). This window functions similarly to the “Approve new variant” window used by the “Declare project variant” function of the Global and Consensus Alignment tabs (see section 1.6.3.3, Figure 1-62, and Figure 2-36, below). The main difference between the use of “Define Haplotype” and “Approve new variant” is a matter of context. The “Approve new variant” function is triggered in the context of a multiple alignment where the linkage of Variants in a haplotype can be visually verified. The “Define Haplotype” function is triggered from the main Variant Tab wherein similar Variant frequencies may be suggestive of a haplotype, but which may also simply be coincidental. Accordingly, the “Define Haplotype” window defaults to creating the haplotype with a “Putative” Status.
If the Variants selected as the haplotype constituents have any contradictory elements (e.g., specifying two different SNPs for the same position or specifying both a SNP and a deletion for the same position), then a window similar to that encountered when the user enters erroneous variant patterns (Figure 1‑29) will be displayed to explain the problem. The user then has the ability to edit the resulting pattern and create a semantically correct haplotype definition; more likely, however, one would select the Cancel button since any identified errors would actually disprove the coincidental existence of the Variants as a valid haplotype.
Figure 1‑52: The “Define Haplotype” window with a Pattern built from the Variant selections made in Figure 1‑51 . The Status defaults to Putative but can be edited, and a red warning is given to indicate that a variant with the same pattern already exists.
1.5.1.5
Editing/Removing Variants from the Variants Tab
Right-clicking on a cell in the “body” of the Variants Frequency Table, at a Sample/Variant intersection (as above in section 1.5.1.3 and as shown in Figure 1-49 G), provides a contextual menu that includes options for editing the Variant Status and for removing Variants from the project. The same functions can be carried out using the Variants Definition Table in the Variants sub-tab of the Project Tab (see section 1.3.2.5). The Variants Frequency Table is a convenient place to decide on the Status of a Variant because the incidence frequency across Samples is available for examination. You can use the shift or control keys to help select multiple rows and you can delete them or adjust their Status as a bulk operation. In the case of removing Variants you are provided with a “Yes to All” confirmation window (similar to the case described in section 1.3.2 and Figure 1‑16). A bulk Status edit occurs without any confirmation step.
1.5.1.6
The Mouse Tracker